Product Parameters
Engineered for precision, built for reliability
- Abbreviation & species
- S.FMDV-O Ab; pig
- Target & method
- Foot-and-mouth disease virus serotype O antibodies; lateral flow immunoassay, visual qualitative reading
- Specimen & loading
- Serum or plasma; add 3 drops (about 90 μL) directly, no buffer
- Reading time
- 5–10 minutes; results after 15 minutes are invalid
- Pack size
- 20 tests per box
- Storage & operation
- Store sealed at 4–30°C away from light, do not freeze; test at 15–30°C
- Shelf Life
- 36 months (when stored unopened at 4–30°C) / Final expiration date is indicated on the package.
- Certifications
- CE, ISO 13485 & GMP certified
- OEM/ODM Service
- Available
Test Procedure
4 steps from the instructions for use
-
1Prepare
Bring cassettes and samples to 15–30°C and lay the cassette flat right after opening.
-
2Draw the sample
Use the disposable dropper provided to draw separated porcine serum or plasma, avoiding air bubbles.
-
3Load the sample directly
Add 3 drops of sample (about 90 μL) vertically into the S well; no buffer is added.
-
4Time and read
Start timing after loading; read and record within 5–10 minutes. Results after 15 minutes are invalid.
Full step-by-step instructions
- 1. Prepare Bring cassettes and samples to 15–30°C, check the packaging and expiry date, and lay the cassette flat on a clean, level surface right after opening.
- 2. Draw the sample Use the disposable dropper provided to draw separated porcine serum or plasma, avoiding air bubbles.
- 3. Load the sample directly Add 3 drops of sample (about 90 μL) vertically into the S well; no buffer is added.
- 4. Time and read Start timing after loading; read and record within 5–10 minutes. Results after 15 minutes are invalid.
Reading the Results
Read within the time stated in the instructions for use

Positive
Both C and T lines appear; a faint colored T line within the reading time is also positive. Interpret with age, vaccination and exposure history.

Negative
Only the C line appears: the antibodies were not detected at the test's detection level; missed detection from early infection or other causes cannot be excluded.

Invalid
No C line appears: invalid whether or not a T line appears. Check the sample, drop count, timing and storage/transport, then retest with a new cassette.
Kit Contents
What is supplied in each box and what you need to prepare

Included in the kit
- Test cassettes
- Disposable droppers
- Instructions for use. This kit has no buffer.
Also required
Timer, suitable blood collection and separation supplies, personal protective equipment