Product Overview
The SaberVet FIV Ab Rapid Test is a highly accurate lateral flow assay intended for the qualitative detection of specific antibodies to FIV. It is an indispensable tool for veterinarians to determine FIV infection status during routine or emergency check-ups.
Product Parameters
Engineered for precision, built for reliability
- Abbreviation & species
- FIV Ab; cat
- Target & method
- FIV-specific antibodies; lateral flow immunoassay, visual qualitative reading
- Specimen
- Whole blood, serum or plasma
- Sample loading
- 1 drop of sample (about 10 μL), then 2 drops of buffer (about 60 μL) into the same well
- Reading time
- 5–10 minutes; results after 15 minutes are invalid
- Pack size
- 10 tests per box
- Storage & operation
- Store sealed at 4–30°C away from light, do not freeze; test at 15–30°C
- Shelf Life
- See the expiration date on the package.
- Certifications
- CE, ISO 13485 & GMP certified
- OEM/ODM Service
- Available
Test Procedure
4 steps from the instructions for use
-
1
Prepare
Bring the kit and sample to 15–30°C and lay the cassette flat right after opening.
-
2Add the sample
Use the capillary dropper provided to add 1 drop (about 10 μL) of whole blood, serum or plasma directly to the S well.
-
3Add the buffer
Immediately add 2 drops of buffer (about 60 μL) to the same S well.
-
4
Time and read
Start timing after loading; read and record within 5–10 minutes. Results after 15 minutes are invalid.
Full step-by-step instructions
- 1. Prepare Bring the cassette, buffer and sample to 15–30°C. Check the product, lot, expiry date and packaging, and lay the cassette flat on a clean, level surface right after opening.
- 2. Add the sample Use the capillary dropper provided to add 1 drop (about 10 μL) of whole blood, serum or plasma directly to the S well.
- 3. Add the buffer Immediately add 2 drops of the buffer provided (about 60 μL) to the same S well. Do not premix the sample or use consumables from other products or lots.
- 4. Time and read Start timing after loading; read and record within 5–10 minutes. Results after 15 minutes are invalid.
Reading the Results
Read within the time stated in the instructions for use

Positive
Both C and T lines appear.

Negative
Only the C line appears: the antibodies were not detected at the test's detection level; with recent exposure or high clinical suspicion, retest or use other methods.

Invalid
No C line appears: invalid with or without a T line. Check the sample, loading, timing and storage/transport, then retest with a new cassette.
Interpretation notes
- Positive: Both C and T lines appear. A faint colored T line within the reading time is also positive; confirm further with the cat's age, vaccination history, exposure and clinical examination.
Performance Data
Method comparison results stated in the instructions for use
| Target | Reference Method | Positive Agreement | Negative Agreement | Overall Agreement |
|---|---|---|---|---|
| FIV Ab | ELISA | 96.88% 155/160 95% CI 92.70%–98.86% | 98.13% 157/160 95% CI 94.38%–99.61% | 97.50% 312/320 95% CI 95.06%–98.81% |
Figures reflect the study samples and reference methods described in the instructions for use. Interpret individual results together with clinical findings.
Kit Contents
What is supplied in each box and what you need to prepare

Included in the kit
- Test cassettes
- Capillary droppers
- Buffer
- Instructions for use
Also required
Timer, suitable blood collection supplies and personal protective equipment