Product Overview
The SaberVet FHV Ag test is specifically optimized for feline practitioners requiring rapid differentiation of respiratory pathogens. By targeting antigens in ocular, nasal, or saliva samples, it delivers a positive agreement of 96.00%. This SaberVet diagnostic tool eliminates the need for expensive lab equipment, providing clear results in 10 minutes. It is an ideal solution for cat shelters and clinics to facilitate immediate isolation and treatment. Each kit is CE certified, ensuring international quality standards for professional veterinary diagnostics.
Product Parameters
Engineered for precision, built for reliability
- Abbreviation & species
- FHV Ag; cat
- Target & method
- Feline herpesvirus antigen; lateral flow immunoassay, visual qualitative reading
- Specimen
- Ocular discharge, nasal discharge or saliva/oral secretions
- Extraction & loading
- Rotate the swab in buffer more than 10 times; 3 drops of mixture (about 90 μL) into the S well
- Reading time
- 5–10 minutes; results after 15 minutes are invalid
- Pack size
- 10 tests per box
- Storage & operation
- Store sealed at 4–30°C away from light, do not freeze; test at 15–30°C
- Shelf Life
- See the expiration date on the package.
- Certifications
- CE, ISO 13485 & GMP certified
- OEM/ODM Service
- Available
Test Procedure
5 steps from the instructions for use
-
1Prepare the buffer
Check the product, lot, expiry date and packaging, and work at 15–30°C. Open the buffer tube provided and place it securely in the workstation.
-
2Collect secretions
Touch the swab to secretions at the ocular, nasal or oral site and note the site.
-
3Extract the sample
Place the swab in the buffer and rotate more than 10 times to release the sample, then close the cap tightly.
-
4Load and time
Lay the cassette flat, add 3 drops of the mixture (about 90 μL) to the S well and start timing.
-
5
Read and record
Read the C/T lines within 5–10 minutes and record the result with a photo of the whole cassette. Results after 15 minutes are invalid.
Full step-by-step instructions
- 1. Prepare the buffer Check the product, lot, expiry date and packaging, and work at 15–30°C. Open the buffer tube provided and place it securely in the workstation.
- 2. Collect secretions Touch the swab provided to secretions at the chosen ocular, nasal or oral site, noting the site and side; avoid wiping only skin or hair.
- 3. Extract the sample Place the swab in the buffer and rotate more than 10 times to release the sample, then close the cap tightly.
- 4. Load and time Lay the cassette flat right after opening, add 3 drops of the mixed sample (about 90 μL) vertically into the S well and start timing. Do not add water, substitute the buffer or keep adding drops.
- 5. Read and record Read the C/T lines within 5–10 minutes and record the result with a photo of the whole cassette. Results after 15 minutes are invalid.
Reading the Results
Read within the time stated in the instructions for use

Positive
Both colored C and T lines appear; a faint T line within the reading time is also positive.

Negative
Only the C line appears: the antigen was not detected at the test's detection level; latent infection, intermittent low-level shedding or sampling differences cannot be excluded.

Invalid
No C line appears: invalid whether or not a line appears in the T zone.
Interpretation notes
- Positive: Both colored C and T lines appear; a faint T line within the reading time is also positive. The antigen was detected in this sample; this does not indicate viral load or prove FHV is the only cause of all signs.
- Invalid: No C line appears: invalid whether or not a line appears in the T zone. Check the sample, extraction, loading and storage/transport, then retest with a new cassette.
Performance Data
Method comparison results stated in the instructions for use
| Target | Reference Method | Positive Agreement | Negative Agreement | Overall Agreement |
|---|---|---|---|---|
| FHV Ag | PCR | 96.00% 192/200 95% CI 92.18%–98.09% | 98.33% 295/300 95% CI 96.04%–99.40% | 97.40% 487/500 95% CI 95.56%–98.52% |
Figures reflect the study samples and reference methods described in the instructions for use. Interpret individual results together with clinical findings.
Kit Contents
What is supplied in each box and what you need to prepare

Included in the kit
- Test cassettes
- Sampling swabs
- Buffer tubes
- Instructions for use
- Workstation
Also required
Timer, sample labels and personal protective equipment