Sabervet Canine Heartworm Antigen/Ehrlichia/Anaplasma/Babesia Gibsoni/Babesiosis Antibody Combo Rapid Test
Product Parameters
Engineered for precision, built for reliability
- Abbreviation & species
- CHW Ag/EHR/ANA/BG/BC Ab; dog
- Antigen target
- CHW: canine heartworm antigen
- Antibody targets
- EHR: Ehrlichia; ANA: Anaplasma; BG: Babesia gibsoni; BC: Babesia canis
- Method & specimen
- Colloidal gold immunochromatography, visual qualitative reading; canine whole blood, serum or plasma
- Sampling & loading
- About 75 μL of blood mixed in buffer; 3 drops of mixture (about 90 μL) into each of the five S wells
- Pack size
- 10 tests per box (five targets per test)
- Reading time
- 5–10 minutes; results after 15 minutes are invalid
- Storage & operation
- Store sealed at 4–30°C away from light, do not freeze; test at 15–30°C
Clear Results, Confident Diagnosis
Eliminate diagnostic uncertainty with our advanced colloidal gold technology that delivers crystal-clear test lines and reliable results every time.
SaberVet Advantage
- Deep red, crystal-clear T-lines
- Clean, interference-free background
- GMP-controlled batch consistency
- Confident, reliable diagnosis
Test Procedure
5 steps from the instructions for use
-
1Prepare the buffer
Open the buffer tube provided and place it in the workstation.
-
2Draw a measured sample
Use the volumetric dropper to draw blood and transfer about 75 μL into the buffer.
-
3Mix
Close the cap and shake the extraction tube gently to mix the blood with the buffer.
-
4Load all five wells
Add 3 drops of the mixture (about 90 μL) to each of the CHW, EHR, ANA, BG and BC wells.
-
5
Time and read
Start timing after loading and read the five windows within 5–10 minutes. Results after 15 minutes are invalid.
Full step-by-step instructions
- 1. Prepare the buffer Open the buffer tube provided and place it in the workstation.
- 2. Draw a measured sample Squeeze the upper bulb of the volumetric dropper to draw blood until the lower measuring tube is full and a little enters the lower bulb; make sure the tube has no air bubbles. Squeeze the upper bulb again to add about 75 μL of blood from the tube into the buffer.
- 3. Mix Close the cap and shake the extraction tube gently to mix the blood with the buffer.
- 4. Load all five wells Lay the cassette on a clean, level surface and add 3 drops of the mixture (about 90 μL) vertically into each of the CHW, EHR, ANA, BG and BC S wells. Do not add whole blood directly, and do not split 3 drops across the five wells.
- 5. Time and read Start timing after loading and read the five windows within 5–10 minutes; note separate start times if wells were loaded far apart. Results after 15 minutes are invalid.
Reading the Results
Read within the time stated in the instructions for use
Check the C/T lines window by window and record CHW Ag, EHR Ab, ANA Ab, BG Ab and BC Ab separately, noting the difference between antigen and antibody results.

Positive
Both C and T lines appear; a faint colored T line within the reading time is also positive.

Negative
Only the C line appears: the corresponding antigen or antibody was not detected at the test's detection level; early infection and other factors can still give a negative result.

Invalid
No C line appears: that item is invalid and is not recorded as negative.
Interpretation notes
- Positive: Both C and T lines appear; a faint colored T line within the reading time is also positive. CHW indicates antigen detected; the other four indicate the corresponding antibodies, and an antibody positive alone does not establish active disease.
- Invalid: No C line appears: that item is invalid and is not recorded as negative. Check the sample, volume and procedure and retest with a new cassette; if the shared sample or buffer was affected, retest all affected targets.
Performance Data
Method comparison results stated in the instructions for use
| Target | Reference Method | Positive Agreement | Negative Agreement | Overall Agreement |
|---|---|---|---|---|
| CHW Ag | PCR | 98.00% 147/150 | 97.50% 156/160 | 97.74% 303/310 |
| EHR Ab | ELISA | 98.50% 197/200 | 97.33% 146/150 | 98.00% 343/350 |
| ANA Ab | ELISA | 98.13% 157/160 | 95.63% 153/160 | 96.88% 310/320 |
| BG Ab | ELISA | 98.00% 147/150 | 97.33% 146/150 | 97.67% 293/300 |
| BC Ab | ELISA | 98.00% 49/50 | 96.67% 145/150 | 97.00% 194/200 |
Each target is evaluated separately against its reference method; the results are not combined into a single accuracy figure for the whole test.
Figures reflect the study samples and reference methods described in the instructions for use. Interpret individual results together with clinical findings.
Kit Contents
What is supplied in each box and what you need to prepare

Included in the kit
- Five-target cassettes
- Volumetric droppers
- Buffer tubes
- Instructions for use
- Workstation
Also required
Timer, blood collection supplies and personal protective equipment