Sabervet Canine Parvovirus/Corona Virus Antigen Combo Rapid Test
Product Overview
This combo rapid test is designed for the qualitative detection of CPV and CCV antigens in canine fecal samples. With positive agreement of 98.00% for CPV and 98.50% for CCV, it provides a reliable tool for diagnosing common canine enteric viruses. The test is user-friendly, with clear results available within minutes, aiding in rapid diagnosis and infection control.
Product Parameters
Engineered for precision, built for reliability
- Abbreviation
- CPV/CCV Ag
- Species
- Dog
- Targets
- Canine parvovirus antigen, canine enteric coronavirus antigen
- Format
- Two-target immunochromatographic combo cassette; wells loaded and read separately
- Specimen
- Fresh feces or rectal swab
- Pack size
- 10 tests per box (two targets per sample)
- Sample volume
- From the same extract, 3 drops (about 90 μL) into each matching S well
- Reading time
- 5–10 minutes; results after 15 minutes are invalid
- Storage & operating temperature
- Store sealed at 4–30°C away from light, do not freeze; test at 15–30°C
- Shelf Life
- See the expiration date on the package.
- Certifications
- CE, ISO 13485 & GMP certified
- OEM/ODM Service
- Available
Test Procedure
5 steps from the instructions for use
-
1Prepare the buffer
Open the buffer tube and place it in the workstation.
-
2Collect the sample
Swab a small amount of fresh feces or collect a rectal sample, then place the swab in the buffer.
-
3Extract the sample
Rotate the swab in the tube more than 10 times to release the sample, then close the dropper cap and mix.
-
4Load each well
Add 3 drops of the mixture (about 90 μL) to each of the CPV and CCV wells.
-
5
Time and read
Start timing as soon as both wells are loaded; read and record each within 5–10 minutes. Results after 15 minutes are invalid.
Full step-by-step instructions
- 1. Prepare the buffer Open the buffer tube and place it in the workstation.
- 2. Collect the sample Use the swab provided to take a small amount of fresh feces, or have trained staff collect a rectal sample, then place the swab in the buffer.
- 3. Extract the sample Rotate the swab in the tube more than 10 times to release the sample, then close the dropper cap and mix.
- 4. Load each well Lay the cassette on a clean, level surface. Add the mixture vertically into the CPV and CCV S wells, 3 drops (about 90 μL) each, not 3 drops in total.
- 5. Time and read Start timing as soon as both wells are loaded; read and record each within 5–10 minutes. Results after 15 minutes are invalid.
Reading the Results
Read within the time stated in the instructions for use
The line rules below apply to the CPV and CCV items separately, as labeled on the cassette.

Positive
Both C and T lines appear for that item; a faint T line also counts.

Negative
Only the C line appears for that item: the corresponding antigen was not detected at the test's detection level, but infection cannot be fully excluded.

Invalid
No C line for that item: invalid whether or not a T line appears.
| CPV | CCV | How to record and interpret |
|---|---|---|
| Positive | Negative | Record both; assess clinically for CPV. |
| Negative | Positive | Note CCV; do not dismiss a strongly suspected case because CPV is negative. |
| Positive | Positive | Both targets detected; assess co-infection and other causes clinically. |
| Negative | Negative | Consider the sample, disease course and signs when deciding on further infectious or non-infectious work-up. |
| Either item invalid | — | Photograph both windows, note the invalid item, investigate and repeat the whole test. |
Interpretation notes
- Invalid: No C line for that item: invalid whether or not a T line appears. The other item's C line cannot substitute; check for missed loading, drop count or a blocked well, then repeat the whole test with a new cassette.
Performance Data
Method comparison results stated in the instructions for use
| Target | Reference Method | Positive Agreement | Negative Agreement | Overall Agreement |
|---|---|---|---|---|
| CPV | PCR | 98.00% 147/150 | 98.67% 148/150 | 98.33% 295/300 |
| CCV | PCR | 98.50% 197/200 | 97.33% 146/150 | 98.00% 343/350 |
Each target is evaluated separately against its reference method; the results are not combined into a single accuracy figure for the whole test.
Figures reflect the study samples and reference methods described in the instructions for use. Interpret individual results together with clinical findings.
Kit Contents
What is supplied in each box and what you need to prepare

Included in the kit
- Dual-test cassettes
- Sampling swabs
- Buffer tubes
- Instructions for use
- Workstation
Also required
Timer