Gives Ehrlichia, Babesia gibsoni and Anaplasma antibody results from one canine whole blood, serum or plasma sample. The buffered sample is loaded into three S wells and each window read within 5–10 minutes, for initial screening when all three infections are being considered.
Specimen
Whole blood, serum or plasma
Pack Size
10 tests per box (three targets per dog)
Reading Time
5–10 minutes
Technical Data

Product Parameters

Engineered for precision, built for reliability

Abbreviation & species
EHR/BG/ANA Ab; dog
Targets
Ehrlichia, Babesia gibsoni and Anaplasma antibodies
Format
Three-target colloidal gold immunochromatographic combo cassette, qualitative
Specimen & processing
Whole blood, serum or plasma; about 75 μL of blood mixed into the buffer provided
Sample volume
3 drops of mixture (about 90 μL) into each of the EHR, BG and ANA S wells
Pack size
10 tests per box (three targets per dog)
Reading time
5–10 minutes; results after 15 minutes are invalid
Storage & operation
Store sealed at 4–30°C away from light, do not freeze; test at 15–30°C

Test Procedure

5 steps from the instructions for use

  1. Step 1: Prepare the buffer
    1

    Prepare the buffer

    Open one buffer tube provided and place it in the workstation.

  2. Step 2: Draw a measured sample
    2

    Draw a measured sample

    Use the volumetric dropper to draw blood and transfer about 75 μL into the buffer tube.

  3. Step 3: Mix
    3

    Mix

    Close the cap tightly and shake gently to mix the sample with the buffer.

  4. Step 4: Load all three wells
    4

    Load all three wells

    Add 3 drops of the mixture (about 90 μL) to each of the EHR, BG and ANA wells.

  5. 5

    Time and read

    Start timing as soon as all three wells are loaded and read the three windows within 5–10 minutes. Results after 15 minutes are invalid.

Full step-by-step instructions
  1. 1. Prepare the buffer Open one buffer tube provided and place it in the workstation.
  2. 2. Draw a measured sample Squeeze the upper bulb of the volumetric dropper to draw blood so that a little sample enters the lower bulb and the tip contains no air bubbles. Squeeze the upper bulb again to transfer about 75 μL of sample into the buffer tube.
  3. 3. Mix Close the cap tightly and shake gently to mix the sample with the buffer.
  4. 4. Load all three wells Lay the cassette on a clean, level surface. Add the same mixture vertically into the EHR, BG and ANA S wells, 3 drops (about 90 μL) per well.
  5. 5. Time and read Start timing as soon as all three wells are loaded and read the three windows within 5–10 minutes. Results after 15 minutes are invalid.

Reading the Results

Read within the time stated in the instructions for use

Each window must show its own C line; record the EHR Ab, BG Ab and ANA Ab results separately.

Positive result

Positive

Both C and T lines appear; a faint colored T line within the reading time also counts as positive, indicating the corresponding antibody was detected.

Negative result

Negative

Only the C line appears: the corresponding antibody was not detected at the test's detection level. When clinical suspicion is high, further testing is needed.

Invalid result

Invalid

No C line appears: the result is invalid whether or not a T line appears.

Interpretation notes
  • Invalid: No C line appears: the result is invalid whether or not a T line appears. Check the sample and procedure and retest with a new cassette; record the other windows by their own lines.

Performance Data

Method comparison results stated in the instructions for use

Target Reference Method Positive AgreementNegative AgreementOverall Agreement
EHR Ab ELISA 98.50% 197/200 97.33% 146/150 98.00% 343/350
BG Ab ELISA 98.00% 147/150 97.33% 146/150 97.67% 293/300
ANA Ab ELISA 98.13% 157/160 95.63% 153/160 96.88% 310/320

Each target is evaluated separately against its reference method; the results are not combined into a single accuracy figure for the whole test.

Figures reflect the study samples and reference methods described in the instructions for use. Interpret individual results together with clinical findings.

Kit Contents

What is supplied in each box and what you need to prepare

Sabervet Canine Ehrlichia/Babesia Gibsoni/Anaplasma Antibody Combo Rapid Test kit components

Included in the kit

  • 10 triple-test cassettes
  • 10 volumetric droppers
  • 10 buffer tubes
  • Instructions for use
  • Workstation

Also required

Timer

Request Information