SaberVet Canine Ehrlichia/ Anaplasma/ Babesiosis Antibody Combo Rapid Test
Product Parameters
Engineered for precision, built for reliability
- Abbreviation & species
- EHR/ANA/BC Ab; dog
- Targets
- Ehrlichia, Anaplasma and Babesia canis antibodies
- Format
- Three-target colloidal gold immunochromatographic combo cassette, qualitative
- Specimen
- Canine whole blood, serum or plasma, processed with the volumetric dropper and buffer provided
- Sample volume
- 3 drops of mixture (about 90 μL) into each of the EHR, ANA and BC S wells
- Pack size
- 10 tests per box (three targets per dog)
- Reading time
- 5–10 minutes; results after 15 minutes are invalid
- Storage & operation
- Store sealed at 4–30°C away from light, do not freeze; test at 15–30°C
Clear Results, Confident Diagnosis
Eliminate diagnostic uncertainty with our advanced colloidal gold technology that delivers crystal-clear test lines and reliable results every time.
SaberVet Advantage
- Deep red, crystal-clear T-lines
- Clean, interference-free background
- GMP-controlled batch consistency
- Confident, reliable diagnosis
Test Procedure
5 steps from the instructions for use
-
1Prepare the buffer
Open one buffer tube provided and place it in the workstation.
-
2Draw a measured sample
Use the volumetric dropper to draw blood and transfer the measured sample into the buffer tube.
-
3Mix
Close the cap tightly and shake gently to mix the sample with the buffer.
-
4Load all three wells
Add 3 drops of the mixture (about 90 μL) to each of the EHR, ANA and BC wells.
-
5
Time and read
Start timing as soon as all three wells are loaded and read the three windows within 5–10 minutes. Results after 15 minutes are invalid.
Full step-by-step instructions
- 1. Prepare the buffer Open one buffer tube provided and place it in the workstation.
- 2. Draw a measured sample Squeeze the upper bulb of the volumetric dropper to draw blood so that a little sample enters the lower bulb and the tip contains no air bubbles. Squeeze the upper bulb again to transfer the sample in the tip into the buffer tube.
- 3. Mix Close the cap tightly and shake gently to mix the sample with the buffer.
- 4. Load all three wells Lay the cassette on a clean, level surface. Add the mixture vertically into the EHR, ANA and BC S wells, 3 drops (about 90 μL) per well.
- 5. Time and read Start timing as soon as all three wells are loaded and read the three windows within 5–10 minutes. Results after 15 minutes are invalid.
Reading the Results
Read within the time stated in the instructions for use
Check the C line in each of the three windows and record the EHR Ab, ANA Ab and BC Ab results separately.

Positive
Both C and T lines appear. A faint colored T line within the reading time also counts as positive, indicating the corresponding antibody was detected.

Negative
Only the C line appears: the corresponding antibody was not detected at the test's detection level. When clinical suspicion is high, further testing is needed.

Invalid
No C line appears: the result is invalid whether or not a T line appears.
Interpretation notes
- Invalid: No C line appears: the result is invalid whether or not a T line appears. Check the sample and procedure and retest with a new cassette; record the other windows by their own lines.
Performance Data
Method comparison results stated in the instructions for use
| Target | Reference Method | Positive Agreement | Negative Agreement | Overall Agreement |
|---|---|---|---|---|
| EHR Ab | ELISA | 98.50% 197/200 | 97.33% 146/150 | 98.00% 343/350 |
| ANA Ab | ELISA | 98.13% 157/160 | 95.63% 153/160 | 96.88% 310/320 |
| BC Ab | ELISA | 98.00% 49/50 | 96.67% 145/150 | 97.00% 194/200 |
Each target is evaluated separately against its reference method; the results are not combined into a single accuracy figure for the whole test.
Figures reflect the study samples and reference methods described in the instructions for use. Interpret individual results together with clinical findings.
Kit Contents
What is supplied in each box and what you need to prepare

Included in the kit
- 10 triple-test cassettes
- 10 volumetric droppers
- 10 buffer tubes
- Instructions for use
- Workstation
Also required
Timer