Sabervet Canine Distemper Virus/Parvovirus Antigen Combo Rapid Test

The Sabervet CDV+CPV Antigen Combo Test detects CDV in eye/nasal/saliva samples and CPV in fecal samples. It offers high accuracy with results in 5–10 minutes, supporting efficient diagnosis of two common canine infections in one kit.
Specimen
Secretions (CDV); feces or rectal swab (CPV)
Pack Size
10 tests per box (two targets per dog)
Reading Time
5–10 minutes
Shelf Life
See the expiration date on the package.

Product Overview

This combo rapid test is designed for the qualitative detection of CDV antigen (from eye, nasal, or saliva samples) and CPV antigen (from fresh feces or rectal swabs). With positive agreement of 96.67% for CDV and 98.00% for CPV, it provides veterinary professionals with a reliable, all-in-one diagnostic solution. The test is user-friendly, with clear visual results available within 10 minutes, aiding in timely clinical decisions.

Technical Data

Product Parameters

Engineered for precision, built for reliability

Abbreviation & species
CDV/CPV Ag; dog
Targets
Canine distemper virus antigen, canine parvovirus antigen
Format
Two-target immunochromatographic combo cassette, qualitative
Specimen
CDV: ocular discharge, nasal discharge or saliva; CPV: fresh feces or rectal swab
Pack size
10 tests per box (two targets per dog)
Sample volume
Each extract into its matching S well, 3 drops (about 90 μL) per well
Reading time
5–10 minutes; results after 15 minutes are invalid
Storage & operation
Store sealed at 4–30°C away from light, do not freeze; test at 15–30°C
Shelf Life
See the expiration date on the package.
Certifications
CE, ISO 13485 & GMP certified
OEM/ODM Service
Available

Test Procedure

6 steps from the instructions for use

  1. Step 1: Prepare two sets of consumables
    1

    Prepare two sets of consumables

    Label the two buffer tubes CDV and CPV, open them and place them in the workstation, with a separate swab for each tube.

  2. Step 2: Collect the CDV sample
    2

    Collect the CDV sample

    Use the CDV swab to collect ocular discharge, nasal discharge or saliva, and place it in the CDV buffer tube.

  3. Step 3: Collect the CPV sample
    3

    Collect the CPV sample

    Use the other swab to collect fresh feces, or have trained staff take a rectal swab, and place it in the CPV buffer tube.

  4. Step 4: Extract each sample
    4

    Extract each sample

    Rotate each swab in its own tube more than 10 times, then close each tube. Keep the two samples separate.

  5. Step 5: Load the matching wells
    5

    Load the matching wells

    Add 3 drops (about 90 μL) from the CDV tube to the CDV well and from the CPV tube to the CPV well.

  6. 6

    Time and read

    Start timing as soon as both wells are loaded and read the two windows within 5–10 minutes. Results after 15 minutes are invalid.

Full step-by-step instructions
  1. 1. Prepare two sets of consumables Label the two buffer tubes CDV and CPV, open them and place them in the workstation, with a separate swab for each tube.
  2. 2. Collect the CDV sample Use the CDV swab to collect ocular discharge, nasal discharge or saliva, and place it in the CDV buffer tube.
  3. 3. Collect the CPV sample Use the other swab to collect fresh feces, or have trained staff take a rectal swab, and place it in the CPV buffer tube.
  4. 4. Extract each sample Rotate each swab in its own tube more than 10 times to release the sample, then close each tube with its dropper cap. Do not mix the two samples.
  5. 5. Load the matching wells Lay the cassette on a clean, level surface. Add 3 drops from the CDV tube vertically into the CDV S well and 3 drops from the CPV tube into the CPV S well, about 90 μL per well. Check the target names on the cassette before loading each well.
  6. 6. Time and read Start timing as soon as both wells are loaded and read the two windows within 5–10 minutes. Results after 15 minutes are invalid.

Reading the Results

Read within the time stated in the instructions for use

Check the C line in each window and record the CDV Ag and CPV Ag results separately.

Positive result

Positive

Both C and T lines appear; a faint T line within the reading time also counts as positive, indicating the corresponding antigen was detected in the sample.

Negative result

Negative

Only the C line appears: the corresponding antigen was not detected at the test's detection level. If infection is strongly suspected, further testing is needed.

Invalid result

Invalid

No C line appears: invalid whether or not a T line appears. Check the sample, drop volume and procedure, then retest with a new cassette.

Performance Data

Method comparison results stated in the instructions for use

Target Reference Method Positive AgreementNegative AgreementOverall Agreement
CDV Ag PCR 96.67% 145/150 97.33% 146/150 97.00% 291/300
CPV Ag PCR 98.00% 147/150 98.67% 148/150 98.33% 295/300

Each target is evaluated separately against its reference method; the results are not combined into a single accuracy figure for the whole test.

Figures reflect the study samples and reference methods described in the instructions for use. Interpret individual results together with clinical findings.

Kit Contents

What is supplied in each box and what you need to prepare

Sabervet Canine Distemper Virus/Parvovirus Antigen Combo Rapid Test kit components

Included in the kit

  • 10 dual-test cassettes
  • 20 sampling swabs
  • 20 buffer tubes
  • Instructions for use
  • Workstation