Sabervet Canine Distemper Virus/Parvovirus Antigen Combo Rapid Test
Product Overview
This combo rapid test is designed for the qualitative detection of CDV antigen (from eye, nasal, or saliva samples) and CPV antigen (from fresh feces or rectal swabs). With positive agreement of 96.67% for CDV and 98.00% for CPV, it provides veterinary professionals with a reliable, all-in-one diagnostic solution. The test is user-friendly, with clear visual results available within 10 minutes, aiding in timely clinical decisions.
Product Parameters
Engineered for precision, built for reliability
- Abbreviation & species
- CDV/CPV Ag; dog
- Targets
- Canine distemper virus antigen, canine parvovirus antigen
- Format
- Two-target immunochromatographic combo cassette, qualitative
- Specimen
- CDV: ocular discharge, nasal discharge or saliva; CPV: fresh feces or rectal swab
- Pack size
- 10 tests per box (two targets per dog)
- Sample volume
- Each extract into its matching S well, 3 drops (about 90 μL) per well
- Reading time
- 5–10 minutes; results after 15 minutes are invalid
- Storage & operation
- Store sealed at 4–30°C away from light, do not freeze; test at 15–30°C
- Shelf Life
- See the expiration date on the package.
- Certifications
- CE, ISO 13485 & GMP certified
- OEM/ODM Service
- Available
Test Procedure
6 steps from the instructions for use
-
1Prepare two sets of consumables
Label the two buffer tubes CDV and CPV, open them and place them in the workstation, with a separate swab for each tube.
-
2Collect the CDV sample
Use the CDV swab to collect ocular discharge, nasal discharge or saliva, and place it in the CDV buffer tube.
-
3Collect the CPV sample
Use the other swab to collect fresh feces, or have trained staff take a rectal swab, and place it in the CPV buffer tube.
-
4Extract each sample
Rotate each swab in its own tube more than 10 times, then close each tube. Keep the two samples separate.
-
5Load the matching wells
Add 3 drops (about 90 μL) from the CDV tube to the CDV well and from the CPV tube to the CPV well.
-
6
Time and read
Start timing as soon as both wells are loaded and read the two windows within 5–10 minutes. Results after 15 minutes are invalid.
Full step-by-step instructions
- 1. Prepare two sets of consumables Label the two buffer tubes CDV and CPV, open them and place them in the workstation, with a separate swab for each tube.
- 2. Collect the CDV sample Use the CDV swab to collect ocular discharge, nasal discharge or saliva, and place it in the CDV buffer tube.
- 3. Collect the CPV sample Use the other swab to collect fresh feces, or have trained staff take a rectal swab, and place it in the CPV buffer tube.
- 4. Extract each sample Rotate each swab in its own tube more than 10 times to release the sample, then close each tube with its dropper cap. Do not mix the two samples.
- 5. Load the matching wells Lay the cassette on a clean, level surface. Add 3 drops from the CDV tube vertically into the CDV S well and 3 drops from the CPV tube into the CPV S well, about 90 μL per well. Check the target names on the cassette before loading each well.
- 6. Time and read Start timing as soon as both wells are loaded and read the two windows within 5–10 minutes. Results after 15 minutes are invalid.
Reading the Results
Read within the time stated in the instructions for use
Check the C line in each window and record the CDV Ag and CPV Ag results separately.

Positive
Both C and T lines appear; a faint T line within the reading time also counts as positive, indicating the corresponding antigen was detected in the sample.

Negative
Only the C line appears: the corresponding antigen was not detected at the test's detection level. If infection is strongly suspected, further testing is needed.

Invalid
No C line appears: invalid whether or not a T line appears. Check the sample, drop volume and procedure, then retest with a new cassette.
Performance Data
Method comparison results stated in the instructions for use
| Target | Reference Method | Positive Agreement | Negative Agreement | Overall Agreement |
|---|---|---|---|---|
| CDV Ag | PCR | 96.67% 145/150 | 97.33% 146/150 | 97.00% 291/300 |
| CPV Ag | PCR | 98.00% 147/150 | 98.67% 148/150 | 98.33% 295/300 |
Each target is evaluated separately against its reference method; the results are not combined into a single accuracy figure for the whole test.
Figures reflect the study samples and reference methods described in the instructions for use. Interpret individual results together with clinical findings.
Kit Contents
What is supplied in each box and what you need to prepare

Included in the kit
- 10 dual-test cassettes
- 20 sampling swabs
- 20 buffer tubes
- Instructions for use
- Workstation
Also required
Timer