Sabervet BoviD-5 Antigen Test(BCoV/Rota/GIA/Cry/E.Coli K99 Ag)Kit
Product Parameters
Engineered for precision, built for reliability
- Species
- Cattle; calf feces or rectal swabs
- Format
- Five independent antigen test units, visual qualitative reading
- Sample loading
- 3 drops of extract (about 90 μL) into each S well
- Reading time
- 5–10 minutes; results after 15 minutes are invalid
- Pack size
- 20 tests per box (five targets per test)
- Storage & operation
- Store sealed at 4–30°C away from light, do not freeze; test at 15–30°C
Test Procedure
5 steps from the instructions for use
-
1Prepare the extraction tube
Bring cassettes, samples and reagents to 15–30°C and check the packaging and expiry date. Open the buffer tube and place it in the workstation.
-
2Collect the fecal sample
Take fresh feces or a rectal swab, covering about 1/2–2/3 of the swab head. Avoid excess feces, bedding and contaminants.
-
3Extract the sample
Place the swab in the buffer, rotate more than 10 times, mix well and close the cap.
-
4Load each well
Add 3 drops of extract (about 90 μL) to each of the five S wells and start timing.
-
5
Time and read
Record the C/T result of each target as labeled on the cassette within 5–10 minutes; results after 15 minutes are invalid.
Full step-by-step instructions
- 1. Prepare the extraction tube Bring cassettes, samples and reagents to 15–30°C and check the packaging and expiry date. Open the buffer tube and place it in the workstation.
- 2. Collect the fecal sample Take fresh feces or a rectal swab, covering about 1/2–2/3 of the swab head. Avoid excess feces, bedding and contaminants.
- 3. Extract the sample Place the swab in the buffer, rotate more than 10 times, mix well and close the cap.
- 4. Load each well Lay the cassette flat after opening, add 3 drops of extract (about 90 μL) vertically into each of the five S wells, and start timing immediately.
- 5. Time and read Record the C/T result of each target as labeled on the cassette within 5–10 minutes; results after 15 minutes are invalid.
Reading the Results
Read within the time stated in the instructions for use

Positive
Both C and T lines appear in that unit; a faint colored T line is also positive.

Negative
Only the C line appears in that unit. The antigen was not detected at the test's detection level; infection is not fully excluded.

Invalid
No C line in that unit: invalid regardless of the T line.
Interpretation notes
- Positive: Both C and T lines appear in that unit; a faint colored T line is also positive. Interpret each target with the calf's age, course of disease and other tests.
- Invalid: No C line in that unit: invalid regardless of the T line. Other targets are read by their own C/T lines and cannot substitute for this unit's control. Check the procedure and retest the invalid target with a new cassette.
Performance Data
Method comparison results stated in the instructions for use
| Target | Reference Method | Positive Agreement | Negative Agreement | Overall Agreement |
|---|---|---|---|---|
| BCoV | PCR | 96.40% 107/111 95% CI 91.10%–98.59% | 95.37% 103/108 95% CI 89.62%–98.01% | 95.89% 210/219 95% CI 92.38%–97.82% |
| Rota | PCR | 98.85% 86/87 95% CI 93.77%–99.80% | 97.96% 96/98 95% CI 92.86%–99.44% | 98.38% 182/185 95% CI 95.34%–99.45% |
| GIA | PCR | 98.57% 69/70 95% CI 92.34%–99.75% | 99.13% 228/230 95% CI 96.89%–99.76% | 99.00% 297/300 95% CI 97.10%–99.66% |
| Cry | PCR | 94.29% 33/35 95% CI 81.39%–98.42% | 98.28% 57/58 95% CI 90.86%–99.69% | 96.77% 90/93 95% CI 90.94%–98.90% |
| E. coli K99 | PCR | 96.55% 28/29 95% CI 82.82%–99.39% | 98.72% 77/78 95% CI 93.09%–99.77% | 98.13% 105/107 95% CI 93.44%–99.49% |
Each target is evaluated separately against its reference method; the results are not combined into a single accuracy figure for the whole test.
Figures reflect the study samples and reference methods described in the instructions for use. Interpret individual results together with clinical findings.
Kit Contents
What is supplied in each box and what you need to prepare

Included in the kit
- Five-target cassettes
- Sampling swabs
- Buffer tubes
- Instructions for use
- Workstation
Also required
Timer, personal protective equipment and biohazard waste container