Sabervet Avian Influenza Virus H5 Antigen Rapid Test
Product Parameters
Engineered for precision, built for reliability
- Abbreviation & species
- AIV H5 Ag; poultry
- Method
- Immunochromatography, single-target qualitative test
- Specimen
- Throat secretions or cloacal swab
- Sample loading
- Extract the swab in buffer; 3 drops of mixture (about 90 μL) into the S well
- Reading time
- 5–10 minutes; results after 15 minutes are invalid
- Pack size
- 20 tests per box
- Storage & operation
- Store sealed at 4–30°C away from light, do not freeze; test at 15–30°C
- Shelf Life
- See the expiration date on the package.
- Certifications
- CE, ISO 13485 & GMP certified
- OEM/ODM Service
- Available
Test Procedure
5 steps from the instructions for use
-
1Prepare
Bring reagents and samples to 15–30°C, lay the cassette flat, and place the opened buffer tube in the workstation.
-
2Collect the sample
Use the swab provided to collect throat secretions or a cloacal sample, keeping the swab away from other surfaces.
-
3Extract the sample
Rotate the swab in the buffer more than 10 times, then close the cap tightly.
-
4Add the mixture
Add 3 drops of the mixed sample (about 90 μL) vertically into the S well.
-
5
Time and read
Start timing immediately after loading; read and record within 5–10 minutes. Results after 15 minutes are invalid.
Full step-by-step instructions
- 1. Prepare Bring reagents and samples to 15–30°C. Check the expiry date and pouch integrity, and lay the cassette flat on a clean, level surface right after opening. Open the buffer tube and place it in the workstation.
- 2. Collect the sample Use the swab provided to collect throat secretions or a cloacal sample, keeping the swab away from other surfaces.
- 3. Extract the sample Place the swab in the buffer tube and rotate more than 10 times to release the sample; remove the swab and close the cap tightly.
- 4. Add the mixture Add 3 drops of the mixed sample (about 90 μL) vertically into the S well.
- 5. Time and read Start timing immediately after loading; read and record within 5–10 minutes. Results after 15 minutes are invalid.
Reading the Results
Read within the time stated in the instructions for use

Positive
Both C and T lines appear; a faint colored T line within the reading time is also positive, indicating H5 avian influenza virus antigen was detected at the test's detection level.

Negative
Only the C line appears: the target was not detected at the test's detection level; infection cannot be fully excluded.

Invalid
No C line appears: invalid whether or not a T line appears. Check sampling, extraction, loading, timing and storage/transport, then retest with a new cassette.
Performance Data
Method comparison results stated in the instructions for use
| Target | Reference Method | Sensitivity | Specificity | Overall Agreement |
|---|---|---|---|---|
| AIV H5 Ag | PCR | 96.47% 164/170 95% CI 92.51%–98.37% | 97.78% 176/180 95% CI 94.43%–99.13% | 97.14% 340/350 95% CI 94.82%–98.44% |
Figures reflect the study samples and reference methods described in the instructions for use. Interpret individual results together with clinical findings.
Kit Contents
What is supplied in each box and what you need to prepare

Included in the kit
- Test cassettes
- Sampling swabs
- Buffer tubes
- Instructions for use
- Workstation
Also required
Timer, suitable personal protective equipment and sample recording supplies