Sabervet African Swine Fever Virus Antigen Rapid Test
Product Overview
Mitigate the risk of African Swine Fever with the Sabervet ASFV Antigen Rapid Test. This critical diagnostic tool detects ASFV antigen in porcine whole blood, serum, or plasma, enabling rapid on-site screening for early outbreak detection. The kit features a precise quantitative dropper system for accurate sample handling. Empower your biosecurity response with fast, reliable results to implement immediate containment measures and protect swine production assets.
Product Parameters
Engineered for precision, built for reliability
- Abbreviation & species
- ASFV Ag; pig
- Target & method
- ASFV antigen; lateral flow immunoassay, visual qualitative reading
- Specimen
- Whole blood, serum or plasma
- Sampling & loading
- About 75 μL of blood mixed in buffer; 3 drops of mixture (about 90 μL) into the S well
- Reading time
- 5–10 minutes; results after 15 minutes are invalid
- Pack size
- 20 tests per box
- Storage & operation
- Store sealed at 4–30°C away from light, do not freeze; test at 15–30°C
- Shelf Life
- See the expiration date on the package.
- Certifications
- CE, ISO 13485 & GMP certified
- OEM/ODM Service
- Available
Test Procedure
4 steps from the instructions for use
-
1Prepare the buffer
Bring everything to 15–30°C, open the buffer tube and place it in the workstation.
-
2Draw a measured sample
Use the volumetric dropper to draw blood and transfer about 75 μL into the buffer tube.
-
3Mix
Close the cap and shake gently to mix the blood with the buffer. Use separate consumables for each sample.
-
4Load and time
Add 3 drops of the mixture (about 90 μL) to the S well and start timing.
Full step-by-step instructions
- 1. Prepare the buffer Bring cassettes, samples and reagents to 15–30°C. Check the lot, expiry date and packaging, open the buffer tube provided and place it in the workstation.
- 2. Draw a measured sample Squeeze the upper bulb of the volumetric dropper to draw blood so that a little liquid enters the lower bulb and the tube contains no air bubbles. Squeeze the upper bulb again to transfer about 75 μL of blood into the buffer tube.
- 3. Mix Close the cap and shake gently to mix the blood with the buffer. Use separate consumables for each sample.
- 4. Load and time Lay the cassette on a clean, level surface and immediately add 3 drops of the mixture (about 90 μL) vertically into the S well. Start timing after loading, and read and record within 5–10 minutes; results after 15 minutes are invalid.
Reading the Results
Read within the time stated in the instructions for use

Positive
Both C and T lines appear; a faint colored T line within the reading time is also positive.

Negative
Only the C line appears.

Invalid
No C line appears: the result is invalid whether or not a T line appears.
Interpretation notes
- Positive: Both C and T lines appear; a faint colored T line within the reading time is also positive. Arrange further confirmation according to the clinical and epidemiological situation.
- Negative: Only the C line appears. The antigen was not detected at the test's detection level; missed detection due to early stage, low levels or sample factors is not excluded.
- Invalid: No C line appears: the result is invalid whether or not a T line appears. Check the sample, procedure and storage/transport conditions, then retest with a new cassette.
Performance Data
Method comparison results stated in the instructions for use
| Target | Reference Method | Positive Agreement | Negative Agreement | Overall Agreement |
|---|---|---|---|---|
| ASFV Ag | PCR | 98.46% 128/130 95% CI 94.56%–99.58% | 98.61% 355/360 95% CI 96.79%–99.41% | 98.57% 483/490 95% CI 97.08%–99.31% |
Figures reflect the study samples and reference methods described in the instructions for use. Interpret individual results together with clinical findings.
Kit Contents
What is supplied in each box and what you need to prepare

Included in the kit
- Test cassettes
- Volumetric droppers
- Buffer tubes
- Instructions for use
- Workstation
Also required
Timer, suitable blood collection and separation supplies and personal protective equipment