SaberVet African Swine Fever Virus Antibody Rapid Test
Product Parameters
Engineered for precision, built for reliability
- Abbreviation & species
- ASFV Ab; pig
- Target & method
- ASFV-specific antibodies; lateral flow immunoassay, visual qualitative reading
- Specimen
- Serum or plasma
- Sample loading
- 1 drop of sample (about 10 μL), then 2 drops of buffer (about 60 μL) into the same well
- Reading time
- 5–10 minutes; results after 15 minutes are invalid
- Pack size
- 20 tests per box
- Storage & operation
- Store sealed at 4–30°C away from light, do not freeze; test at 15–30°C
Clear Results, Confident Diagnosis
Eliminate diagnostic uncertainty with our advanced colloidal gold technology that delivers crystal-clear test lines and reliable results every time.
SaberVet Advantage
- Deep red, crystal-clear T-lines
- Clean, interference-free background
- GMP-controlled batch consistency
- Confident, reliable diagnosis
Test Procedure
4 steps from the instructions for use
-
1
Prepare
Bring cassettes, reagents and samples to 15–30°C and lay the cassette flat right after opening.
-
2Add the blood sample
Use the capillary dropper provided to draw serum or plasma and add 1 drop (about 10 μL) to the S well.
-
3Add the buffer
Immediately add 2 drops of the buffer provided (about 60 μL) to the same S well.
-
4
Time and read
Start timing after loading; read and record within 5–10 minutes. Results after 15 minutes are invalid.
Full step-by-step instructions
- 1. Prepare Bring cassettes, reagents and samples to 15–30°C. Check the product, lot, expiry date and packaging, and lay the cassette flat on a clean, level surface right after opening.
- 2. Add the blood sample Use the capillary dropper provided to draw serum or plasma and add 1 drop (about 10 μL) to the S well.
- 3. Add the buffer Immediately add 2 drops of the buffer provided (about 60 μL) to the same S well.
- 4. Time and read Start timing after loading; read and record within 5–10 minutes. Results after 15 minutes are invalid.
Reading the Results
Read within the time stated in the instructions for use

Positive
Both C and T lines appear; a faint colored T line within the reading time is also positive. Evaluate further with vaccination, exposure and clinical history.

Negative
Only the C line appears. Antibodies were not detected at the test's detection level; missed detection from early infection or other causes is not excluded.

Invalid
No C line appears: the result is invalid whether or not a T line appears.
Interpretation notes
- Invalid: No C line appears: the result is invalid whether or not a T line appears. Check the sample, procedure and storage/transport conditions, then retest with a new cassette.
Performance Data
Method comparison results stated in the instructions for use
| Target | Reference Method | Positive Agreement | Negative Agreement | Overall Agreement |
|---|---|---|---|---|
| ASFV Ab | ELISA | 99.17% 119/120 95% CI 95.43%–99.85% | 99.17% 357/360 95% CI 97.58%–99.72% | 99.17% 476/480 95% CI 97.88%–99.68% |
Figures reflect the study samples and reference methods described in the instructions for use. Interpret individual results together with clinical findings.
Kit Contents
What is supplied in each box and what you need to prepare

Included in the kit
- Test cassettes
- Capillary droppers
- Buffer
- Instructions for use
Also required
Timer, suitable blood collection and separation supplies, personal protective equipment